# Compact 300 kVp line-focus X-ray tube: microbeam plans for superficial tumours feasible in simulation

- Published: 2026-11-24 05:33 UTC
- Section: Life & Health › Medtech
- Tags: microbeam radiotherapy, compact X-ray tube, line-focus X-ray tube, 300 kVp, treatment planning study, peak-to-valley dose ratio, superficial tumours, equivalent uniform dose, first-in-human trials, Monte Carlo dose calculation
- URL: https://realtech.media/news/rt-d899761252c6645554a65f3b_compact_300_kvp/

**A planning study on 12 retrospectively selected patient cases found dosimetrically feasible microbeam plans for superficial tumours; no patients were treated.**

![Compact 300 kVp line-focus X-ray tube: microbeam plans for superficial tumours feasible in simulation](https://realtech.media/media/2026/11/09b8a313f048c27b.jpg)
*Rich257, Wikimedia Commons, CC BY-SA 4.0*

Medical physicists at the TUM University Hospital Department of Radiation Oncology evaluated microbeam radiotherapy (MRT) planning with a compact line-focus X-ray tube (LFXT). The study, published in Radiation Oncology, is computational: treatment plans were calculated on patient anatomies, and no patients were irradiated. According to the authors, the plans were dosimetrically feasible for superficial tumours [1].

MRT delivers spatially fractionated beams on a micrometre scale. In preclinical studies it spared normal tissue better than conventional radiotherapy while maintaining tumour control, but it has mostly been limited to large synchrotron facilities. The compact LFXT, with a 300 kVp beam and an eccentric focal spot, offers a route to clinical use, and the study aimed to identify suitable scenarios for first-in-human trials [1].

## Patient selection and planning method

Patients were screened retrospectively against clinical criteria: palliative indication, superficial tumour location, reproducible and safe positioning, minimal interference from other therapies, and expected benefit from MRT. From 35 preselected patients, 12 cases were chosen. They included primary lymphoma, squamous cell carcinoma and superficial soft tissue metastases [1].

Three-dimensional forward planning was done in Eclipse 16.1 with two coplanar fields, manually chosen gantry angles and conformal masks. The model assumed 50 μm slits at 400 μm centre-to-centre spacing, with the collimator rotated 90° between fields. Dose was calculated with a Monte Carlo-based hybrid algorithm; average uncertainty was 3–4% for peak and 5–6% for valley doses. A single-fraction equivalent uniform dose of 3 Gy was prescribed to the target volume [1].

## Dosimetric results

Under exploratory assumptions, plans for gluteal, nuchal and inguinal tumours showed optimal target coverage and organ-at-risk sparing. Maximum organ-at-risk doses stayed below 8.5 Gy, except in bone or overlapping high-density regions. Across all plans the mean peak dose was 57.2 Gy, the mean valley dose 3.9 Gy, and the mean peak-to-valley dose ratio 15.4 [1].

The authors observed dose inhomogeneities caused by tissue heterogeneity, kilovoltage beam interactions and a limited range of suitable gantry angles. Tumours in anatomically complex regions, near organs at risk or tissue inhomogeneities, showed less favourable trade-offs between target coverage and organ sparing [1].

The study names superficial tumours as the optimal initial indications. The peak, valley and ratio values agree with preclinical evidence, but their clinical significance remains uncertain, and the authors state that safety-driven dose escalation studies will be required in early-phase trials [1].

## Sources

1. [Radiation Oncology](https://doi.org/10.1186/s13014-026-02946-2)
2. [bioengineer.org](https://bioengineer.org/compact-x-ray-microbeam-treatment-planning-shows-feasibility-for-superficial-tumors)

---

Source: RealTech, realtech.media — https://realtech.media/news/rt-d899761252c6645554a65f3b_compact_300_kvp/