# Paderborn University detector counts 0 to over 9,000 photons, with sub-single-photon precision up to 276

- Published: 2026-11-17 20:50 UTC
- Section: Science › Instruments & experiments
- Tags: photon-number-resolving detector, Poisson noise limit, photon counting, detector multiplexing, quantum detector tomography, optical power standard, Paderborn University, Optica
- URL: https://realtech.media/news/rt-491b87cc282925d62661a45e_paderborn_university_detector/

**Eight superconducting nanowire detectors multiplexed across 128 time modes beat the Poisson noise limit by at least 4.1 dB in a laboratory test with laser pulses.**

![Paderborn University detector counts 0 to over 9,000 photons, with sub-single-photon precision up to 276](https://realtech.media/media/2026/11/12e17668e3e54c9e.jpg)
*NASA/JPL-Caltech, Wikimedia Commons, Public domain*

Researchers at Paderborn University in Germany have demonstrated a detection system that counts photons from 0 to more than 9,000, according to a paper in Optica [1]. Across this range the system beats the Poisson noise limit by at least 4.1 dB [1]. Precision below a single photon was reached up to 276 photons per pulse [1].

## How the detector is built

The device multiplexes eight superconducting nanowire single-photon detectors (SNSPDs), each intrinsically able to resolve photon number, across 128 temporal modes [1]. A single light pulse is divided into 128 smaller pulses separated in time and distributed among the eight detectors, which gives 1,024 detection bins. The authors say low-loss optical components spread the light evenly across the bins while preserving photon-number information in each of them.

Each of the 1,024 bins was characterized with a detector model, previously developed together with Mariia Sidorova of PTB in Germany, to improve measurement precision [1]. The team also performed quantum detector tomography to reconstruct the positive operator-valued measures (POVMs) of the complete device, which consist of 1.38·10^8 matrix elements [1].

## Test results and limits

The researchers tested laser pulses averaging between 1 and 15,000 photons. At the experiment's repetition rate of 80 kHz, the system can precisely count photons corresponding to an optical power of 71 pW [1]. The authors describe this as bridging the gap between single-photon measurements and high-sensitivity optical power meters. According to the authors, a conventional multiplexed system would need about 37,000 detector elements for comparable precision, compared with 1,024 detection bins here.

So far the system has been tested only with laser pulses. The team plans to apply it to bright quantum states of light to investigate quantum effects at larger scales. The authors state that photon-number-resolving detectors of this size, and the tools used to analyze them, will become more important for characterizing large quantum states, precision metrology and optical power standards [1]. They also link the work to photonic quantum computing and single-photon-level imaging.

## Sources

1. [Optica](https://doi.org/10.1364/OPTICA.604389)
2. [SciTechDaily](https://scitechdaily.com/quantum-breakthrough-scientists-develop-a-detector-that-can-count-over-9000-photons)

---

Source: RealTech, realtech.media — https://realtech.media/news/rt-491b87cc282925d62661a45e_paderborn_university_detector/