# Salt-gradient one-step heparin coating keeps 73% of heparin after 16 days in lab tests on polylactic acid

- Published: 2026-11-23 22:49 UTC
- Section: Machines & Making › Materials science
- Tags: anticoagulant coating, heparin coating, salt concentration gradient, liquid–liquid phase separation, polyelectrolyte complex, PEI heparin, thrombosis blood-contacting devices, cardiovascular implants, polylactic acid
- URL: https://realtech.media/news/rt-11cdb1c69f3c6095cebf4ce5_salt_gradient_one/

**Researchers deposited a PEI–heparin anticoagulant coating in one immersion-withdrawal step. In vitro and ex vivo, it cut platelet adhesion by 98% and thrombus formation by 90%.**

![Salt-gradient one-step heparin coating keeps 73% of heparin after 16 days in lab tests on polylactic acid](https://realtech.media/media/2026/11/f3e8a40e99cf0936.jpg)
*Edward Roberts (edward.beardsell)/www.clinicccompare.co.uk, Wikimedia Commons, CC BY 2.0*

Researchers have reported a one-step method for applying anticoagulant coatings to blood-contacting materials, according to a paper in Surface Science and Technology. The coating is built from polyethyleneimine (PEI) and heparin and is driven by a salt-concentration gradient. The tests were laboratory and ex vivo experiments on polylactic acid substrates; the coating has not been reported in clinical use. [1]

Surface-induced thrombosis limits the safety of blood-contacting devices. Surface heparinization is widely used to improve hemocompatibility, but existing methods, such as covalent grafting and Layer-by-Layer assembly, are complex or not stable enough, the authors state. [1]

## How the coating is formed

The team used 4 mol/L NaCl to screen the electrostatic interactions between PEI and heparin, which gave a homogeneous precursor slurry. According to the publisher's account, the polyelectrolyte complexes were first precipitated at acidic pH, centrifuged at 5,000 rpm for 5 minutes and then redissolved in the salt solution. [1]

Polylactic acid specimens were pretreated with dopamine, then with a dopamine and low-molecular-weight PEI solution. A tube was filled with deionized water above a layer of the high-salt slurry, forming a liquid interface. The substrate was immersed in the slurry and withdrawn upward through the interface into the water. [1]

Salt diffuses rapidly into the water at the interface, the charge screening disappears, and liquid–liquid phase separation deposits a dense coacervate of PEI and heparin on the surface. The authors attribute the stability of the coating to physical entanglement of the polymer chains. [1]

Withdrawal speed affected the morphology. At 5 and 10 mm/s the coating did not form a continuous film and instead formed isolated island-like clusters. At 1 mm/s the film was continuous and uniform. [1]

## Measured properties

The coated surface retained 73% of its heparin after 16 days of physiological incubation. The hemolysis rate was 0.32%, platelet adhesion was suppressed by 98%, and ex vivo thrombus formation was reduced by 90%. [1]

The water contact angle fell from about 83.98° on bare polylactic acid to 24.92° on the coated surface. Infrared spectroscopy showed a peak near 937 cm⁻¹ assigned to sulfate ester bonds characteristic of heparin. X-ray photoelectron spectroscopy found nitrogen (14.02 atomic %) and sulfur (2.04 atomic %) on the coating, which were absent on the bare substrate. [1]

The authors describe the approach as a scalable assembly route for functionalizing cardiovascular implants. [1]